Instructions for Use — OtoJig System

only

The latest versions of all instructions for use are available at:
https://otojig.com/IFU/.

Version of this document: v0.4, issued 2026-01-06, OtoJig GmbH. All rights reserved.

Description and Identification of Devices

The OtoJig procedure involves a number of medical devices which, when inter-connected or combined, are intended to achieve the following medical purpose:

The OtoJig system is intended to create a minimally invasive surgical tunnel in the temporal bone and to provide a covered access to the inner ear.

Healthcare Professionals can verify the suitability of the OtoJig devices by consulting this instructions for use. The selection of corresponding software and accessories is limited to the following lists of OtoJig devices and compatible 3rd-party devices.

current-master-presentation_Extract.001

Figure 1: Devices of the OtoJig System (images not to scale; color and appearance not as final devices). The basic working principle is that a Mini-Stereotactic Frame is rigidly bone-anchored for the duration of the surgery. A patient individual Positioning Jig is mounted with Jig Fastener to restrict drill bits and the Round Window opener with a Tool Guide to a previously planned straight access tunnel, thereby allowing a minimally invasive procedure. The Insertion Tube can be used to provide a temporal cover in this drilled access tunnel for cochlear insertions. The X-Ray Marker can be localized in radiological volume image by the planning software conCIerge to establish a registration between the image and patient space. Based on this registration, a virtual access tunnel can be planned and according to this planning, a patient-individualized Positioning Jig is produced on-demand. The Optical Reference Marker is used by conCIerge to assure the Mini-Stereotactic Frame has not been repositioned.

OtoJig System

The OtoJig System includes the following OtoJig and third-party devices listed below. The mutual compatibility of the devices has been verified in accordance with the manufacturer's instructions, and all required activities have been performed in accordance with those instructions.

OtoJig Devices

Device nameManufacturerArticle Number
Mini-Stereotactic Frame (MSF)OtoJig GmbHREF 121.0
Optical Reference Marker (part of the MSF)OtoJig GmbHREF 121.0
AnchorOtoJig GmbHREF 101.0
Mounting ScrewOtoJig GmbHREF 110.0
Jig Fastener (Shell and Pin)OtoJig GmbHREF 150.0
Tool GuideOtoJig GmbHREF 161.1
Positioning JigOtoJig GmbHREF 131.0
Round Window OpenerOtoJig GmbHREF 170.0
Insertion TubeOtoJig GmbHREF 180.0
X-Ray MarkerOtoJig GmbHREF 141.0
conCIerge (planning software)OtoJig GmbHv1.*

Warning

Only combine devices as specified in this IFU. Any other combination of devices is not permitted and can result in injury to the patient or user.

Third-Party Devices

ManufacturerNameDeviceArticle Number(s)
KLS Martin SE & Co. KGBone ScrewmaxDrive®, screw, ⌀ 2.0 mm, length 6 mm25-872-06-09 (pack of 5)
25-872-06-61 (piece)
25-872-06-71 (piece, sterile)
25-872-06-75 (pack of 5, sterile)
KLS Martin SE & Co. KGEmergency Bone ScrewmaxDrive®, screw, ⌀ 2.3 mm, length 6 mm25-873-06-09 (pack of 5)
25-873-06-71 (piece, sterile)
25-873-06-75 (pack of 5, sterile)
KLS Martin SE & Co. KGScrew driver bladeScrew driver blade MaxDrive 2.0/2.3 mm, L25-486-97-07
Gebr. Brasseler GmbH & Co. KGOtoDrill 11Twist drill round shank ⌀ 3.00 mm
S27243.501
REF 191.1
ID 10029330
Gebr. Brasseler GmbH & Co. KGOtoDrill 21Twist drill round shank ⌀ 1.30 mm
S27244.501
REF 192.1
ID 10029331
Gebr. Brasseler GmbH & Co. KGOtoDrill 31Bur, cylindrical round shank ⌀ 1.23 mm
ND0404.013
REF 193.1
ID 10029332
Gebr. Brasseler GmbH & Co. KGPre-drill BitTwist drill round shank ⌀ 1.5 mm
S27245.035
REF 194.0
ID 10029329

Warning

Only use the above listed bone screws in combination with the Anchor (REF 101.0).

Materials, Supplied Condition, and Reusability

The table below lists the materials and substances from which the OtoJig devices are made, the supplied condition, and the reusability. All (hardware) devices are to be used in sterile condition.

DeviceMaterialDirect contact
with tissue
Duration of
body contact
Supplied
condition
Reusability
Mini-Stereotactic Frame (MSF)Titanium alloy Ti6Al4V ISO 5832-3YesShort termnon-sterilereusable (50x)
Optical Reference Marker (part of the MSF)Titanium alloy Ti6Al4V ISO 5832-3YesShort termnon-sterilereusable (50x)
AnchorTitanium alloy Ti6Al4V ISO 5832-3YesShort termnon-sterilesingle use
Mounting ScrewTitanium alloy Ti6Al4V ISO 5832-3NoShort termnon-sterilesingle use
Jig Fastener (Shell and Pin)Titanium alloy Ti6Al4V ISO 5832-3NoShort termnon-sterilesingle use
Tool GuideChromium stainless steel: 1.4112 (EN 10027-2)NoShort termnon-sterilesingle use
Positioning JigPolypropylene PP (ISO 1043-1)NoShort termnon-sterilesingle use
Round Window OpenerChromium stainless steel 1.4112 (EN 10027-2)YesTransientnon-sterilesingle use
Insertion TubePolyimideYesTransientsterile (radiation sterilization)single use
X-Ray MarkerPolyetheretherketone PEEK (ISO 1043-1)NoTransientnon-sterilereusable (50x)
ConCIerge (planning software)SoftwareN/AN/AN/AN/A

Intended Purpose

ConCIerge is intended to plan minimally invasive access tunnels in the temporal bone.

The Mini-Stereotactic Frame (Including Optical Reference Marker) is intended to be attached to the skull behind the patient’s ear by using the Mounting Screw and the Anchor. It is intended to carry the Positioning Jig or the X-Ray Marker.

The Anchor is intended to serve as a fixation point attached to the skull behind the ear.

The Mounting Screw is an accessory for the medical device Mini-Stereotactic Frame. The Mounting Screw is intended to keep the Mini-Stereotactic Frame in position on the skull behind the patient’s ear.

The Jig Fastener is an accessory for the medical devices Positioning Jig and X-Ray Marker. The Jig Fastener is intended to fasten either the Positioning Jig or the X-Ray Marker onto defined mounting sockets.

The Tool Guide is intended to restrict movements of Round Window Opener or drill bits along a pre-planned access tunnel.

The Positioning Jig is intended to position a Tool Guide according to a pre-planned access tunnel relative to the position of the Mini-Stereotactic Frame.

The Round Window Opener is intended to open the round window membrane of the cochlea.

The Insertion Tube is intended to provide a covered canal in the temporal bone.

The X-Ray Marker, whilst mounted on the Mini-Stereotactic Frame, is intended to be locatable by a planning software in a computed tomography volume image.

Medical Indications, Contraindications & Patient Population

All OtoJig devices are intended for use with adult patients.

Caution

Do not use the devices for children.

ConCIerge is indicated for:

ConCIerge is contraindicated for:

The Mini-Stereotactic Frame is indicated for surgical procedures that benefit from the support of the Positioning Jig or the X-Ray Marker in a position attached to the patient’s skull behind the ear.

The Mini-Stereotactic Frame is contraindicated for:

The Anchor is indicated for surgical procedures that benefit from a fixation point attached to the skull behind the ear.

The Anchor is contraindicated for:

The Mounting Screw is an accessory for the Mini-Stereotactic Frame.

The Mounting Screw is contraindicated if the Mini-Stereotactic Frame is contraindicated.

The Jig Fastener is an accessory for the Positioning Jig and the X-Ray Marker.

The Jig Fastener is contraindicated if the Positioning Jig and the X-Ray Marker are contraindicated.

The Tool Guide is indicated for surgical procedures that benefit from restricting movement of the Round Window Opener or drill bits along a pre-planned access tunnel.

The Tool Guide is contraindicated for patients with suspected sensitivity or allergies against chromium stainless steels.

The Positioning Jig is indicated for surgical procedures which benefit from the positioning of a Tool Guide according to a pre-planned access tunnel in relative position to a Mini-Stereotactic Frame.

The Positioning Jig is contraindicated for patients with suspected sensitivity or allergies against Polypropylene (PP).

The Round Window Opener is indicated for surgical procedures that require opening of the round window membrane of the cochlea.

The Round Window Opener is contraindicated for patients with

The Insertion Tube is indicated to form a covered canal in the temporal bone.

The Insertion Tube is contraindicated for patients with

The X-Ray Marker is indicated for surgical procedures that benefit from a marker which is locatable in a computed tomography volume image.

The X-Ray Marker is contraindicated for patients:

Intended User and User Training

Intended user

Training

Training is provided by OtoJig. It includes personal handling of representative devices on a temporal bone model under the supervision of a qualified trainer.

Warnings and Precautions

General safety information

Warning

For safety reasons, please read these Instructions for Use carefully before using the devices. For the device conCIerge, please consult the separate eIFU of conCIerge.

For non-sterile delivered single-use devices: processing must be performed according to the following processing instructions.

For sterile delivered single-use devices: do not use devices beyond the use-by date indicated on the sterile packaging.

For reusable devices: cleaning, disinfection and sterilization must be performed prior to first use and after each use according to the reprocessing instructions. Do not reuse more than 50 times as indicated.

For single-use devices: Reuse and/or reprocessing is prohibited. Reusing and/or reprocessing a single-use device can result in biocontamination, degraded performance, or loss of functionality.

Before use, the packaging of all sterilized devices must be inspected visually in an unaided way to determine if damages in the sterile barrier system integrity are evident. Do not use if the sterile packaging is damaged.

Caution

Do not use the devices for children.

Devices with direct tissue contact are contraindicated if there are suspected sensitivities or allergies to the used materials. See Contraindications.

Before processing of non-sterile delivered devices, the packaging of all non-sterile delivered devices must be inspected visually in an unaided way to determine damages of the packaging. Do not use devices if the packaging is damaged.

All devices must be inspected visually in an unaided way for damage, wear, corrosion, and legibility of the labeling. Damaged, worn, or corroded devices as well as devices with illegible labeling must not be used. Do not use devices more often than the specified reusable cycles.

For single-use devices: discard them after the procedure, even if they have not been used.

The devices and their components must not be modified and must be used only as intended by this instructions for use.

Handle the devices carefully, do not apply excessive force as this may cause mechanical damage to the devices or tissue in direct contact with the devices.

Mini-Stereotactic Frame (incl. Optical Reference Marker)

Warning

Cut hazard: Its sharp spikes can potentially pierce gloves and skin.

Caution

Handle with care. Dispose the device if it might have been bent or deformed (e.g. dropped to the floor).

X-Ray Marker

Caution

Handle with care. Dispose the device if it might have been subjected to a strong impact (e.g. dropped to the floor).

Positioning Jig

Warning

Patient individualized Positioning Jigs are intended to be solely used for the patient for which they have been produced.

Round Window Opener

Warning

The Round Window Opener has a sharp tip, handle with care.

The Round Window Opener must be guided in the Tool Guide.

Caution

In the case of local anesthesia: pay attention to possible movements of the patient's head when using the Round Window Opener. Movements of the patient's head near the tip of the Round Window Opener can lead to injuries.

Insertion Tube

Caution

Devices larger than the lumen of the Insertion Tube cannot be used.

Do not apply large forces on the Insertion Tube as this may damage the device or tissue. Stop inserting the Insertion Tube when resistance is felt or the tube has reached the desired depth.

Residual Risks and Undesirable Side-Effects

The medical devices are used in the context of or as part of other surgically invasive procedures such as cochlear implantation, middle ear surgery, or similar surgeries. The risks associated with the OtoJig procedure inherit the primary risks of those (conventional) surgeries, which contain: excessive bleeding, blood clots, death, delayed wound healing, infections, wound dehiscence, breathing difficulties due to infection, inflammation, diffuse swelling, tissue necrosis, necrosis, hematoma, leakage of cerebrospinal fluid, damage to the facial nerve, pain, scarring at the wound site, exanthema, skin lesions, skin irritation, difficulty swallowing, and complications associated with anesthesia. In addition, meningitis is a rare postoperative complication with cochlear implantation, but can be potentially serious. As with all skin incisions, the potentially additional incisions or puncturing of the skin can often lead to local wound healing disorders, infection, inflammation, or itching. The occurrence probability of these risks is influenced by multiple factors, such as surgical technique, skin incisions and the duration of the surgery.

The device-specific residual risks are:

RiskDevice(s)                   
There is an improbable risk that the planned surgery needs to be aborted or switched to conventional surgical techniques, if there are reasons to believe that the procedure cannot be safely completed, in particular if any of the safety measures trigger an alert, including facial nerve monitoring, intraoperative checklists and other risk controls implemented in software.conCIerge
There is a frequent risk that either no (safe) access tunnel can be planned for a given patient due to anatomic constraints or that no Positioning Jig can be produced for a given planned access tunnel. This results in a non-availability of the minimally-invasive treatment that lead to abort of the surgery or switch to conventional surgical techniques.  conCIerge
There is an occasional risk that the OtoJig system cannot be mounted because of anatomical constraints. In this case, the surgery needs to be aborted or switched to conventional surgical techniques.Anchor, Mounting Screw, Mini-Stereotactic Frame, Positioning Jig, Jig Fastener, Tool Guide
There is an occasional risk of nerve damages related to device deficiencies that can lead to deviations of the drill hole from the planned path. Anchor, Mounting Screw, Mini-Stereotactic Frame, Positioning Jig, Jig Fastener, Tool Guide
There is an occasional risk of nerve damages associated with nonconforming devices (production errors).Positioning Jig
There is an occasional risk of nerve damages associated with drilling inaccuracies resulting from incorrect assembly of the Jig Fasteners.Positioning Jig, X-Ray Marker
There is an occasional risk of temporary or permanent injuries by thermal damage and/or by direct mechanical damage2 resulting from:
- software defects,
- cyber-security incidents (including forging, hacking, data manipulation etc.),
- planning and approval of an incorrect or unfavorable access tunnel,
- compatible drill bits (OtoDrills) deviating from the planned path due to excessive user-applied forces,
- inadequate hard tissue (e.g. inhomogeneous bone densities).
conCIerge
There is an occasional risk of damaging the meninges or a blood vessel, which could result in bleeding or infections, during pre-drilling. conCIerge
There is a frequent risk that device deficiencies (e.g. production errors, device damages) lead to deviations in the axis of the Round Window Opener or prevent fitting of the Round Window Opener into the Tool Guide. A new device may be required.Round Window Opener
There is a frequent risk that device deficiencies (e.g. production errors, device damages) lead to loss of device functionality. A new device may be required.Insertion Tube
There is an occasional risk that device deficiencies (e.g. production errors) lead to loss of device functionality. A new device may be required.Anchor, Mounting Screw, Mini-Stereotactic Frame, Positioning Jig, Jig Fastener, Tool Guide, X-Ray Marker
There is a frequent risk of cuts/pricks (sharp tips) when handling the devices.Round Window Opener
Mini-Stereotactic Frame
There is an occasional infection risk when handling the used devices for disposal.All OtoJig hardware devices
There is the risk when using the Insertion Tube out of the given geometrical limits (inner diameter, outer diameter, length). These risks cannot be predicted in advance.Insertion Tube
There is an occasional risk of tissue damages/cuts when applying high forces to the inserted Insertion Tube.Insertion Tube
There is an occasional risk of damaging/contamination the Insertion Tube during aseptic presentation. A new device may be required.Insertion Tube
There is a very frequent risk of injuries such as cuts/pricks associated with the fixation of the Anchor: If the initial pre-drilled hole is found to be unsuitable, it may be necessary to drill a second one.Anchor
There is a very frequent risk of injuries such as cuts/pricks associated with the fixation of the Mini-Stereotactic Frame: If the Mini-Stereotactic Frame is not properly positioned behind the ear, a second pre-drilled hole might need to be drilled to reposition the device.Mini-Stereotactic Frame
There is an occasional risk of injuries such as cuts/pricks associated with raised edges (burrs).Anchor, Mounting Screw, Mini-Stereotactic Frame, Positioning Jig, Jig Fastener, Tool Guide, X-Ray Marker
There is an occasional risk that the OtoJig system cannot be assembled because of usability issues. In this case, the surgery needs to be aborted or switched to conventional surgical techniques.Anchor, Mounting Screw, Mini-Stereotactic Frame, Positioning Jig, Jig Fastener, Tool Guide
There is an occasional risk of fingers/skin becoming trapped when using the device.Jig Fastener
There is an occasional risk of OtoDrills getting stuck in the Tool Guide. New devices may be required.Tool Guide
There is an occasional risk of delays during surgery: Depending on the level of experience of the user or on the occurrence of software anomalies, the intraoperative planning may take longer than usual and thereby cause negative effects from a prolonged anesthesia duration.conCIerge
There is an improbable risk of non-availability of treatment due to: iPad device failures or the device running out of battery, or software anomalies, or if the X-Ray Marker cannot be localized in the imported CT volume image.conCIerge
There are improbable risks related to privacy disclosures resulting from to unauthorized access to user, patient, and/or (medical) planning data or manipulation of the software or data.conCIerge
There is a very rare risk of allergic reactions against the materials used (chrome-containing stainless steel, titanium, polyimide, polypropylene, and polyetheretherketone).All OtoJig hardware devices

Residual risks are described according to the following probabilities per use:

Note

Off-label use, misuse, combination with non-compatible devices, modification of devices or contraindications may lead to additional risks. Depending on the type of misuse, there may be additional risks that cannot be predicted in advance.

Expected Clinical Benefits

The use of the OtoJig procedure offers several advantages, including reduced surgical drilling time and less healthy bone volume removal compared to traditional mastoidectomy procedures:

General Processing Instructions

Devices: Anchor, Mounting Screw, Jig Fastener, Tool Guide and Round Window Opener

Processing instructions according to ISO 17664-1.

Warning

It is the responsibility of the user to use the validated processing recommended in these instructions for use and to use suitable equipment qualified and validated according to ISO 15883 series and other national recommendations.

The instructions provided in these instructions for use have been validated by the manufacturer of the non sterile single-use medical devices as being capable of preparing a medical device for use. It remains the responsibility of the processor to ensure that the processing, as actually performed, using equipment, materials, and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.

1 Processing

1.1 Automated Cleaning, Disinfection and Drying

Perform automated cleaning, disinfection and drying:

  1. Prewash with cold water for at least 4 minutes – drain

  2. Wash with a detergent (Neodisher® MediClean forte (1%)) at +55 °C for at least 5 minutes – drain

  3. Rinse with cold water3 for at least 1 minute – drain

  4. Disinfect with hot water3 at +93 °C for at least 5 minutes – drain

  5. Dry with hot air at +100 °C for at least 20 minutes

Validation was performed with a washer/disinfector Belimed WD 290 IQ.

1.2 Inspection

Visually inspect the device for damages, remaining contaminations, and moisture. If there are contaminations or moisture, repeat cleaning/disinfection process and inspection.

Caution

Discard the device if damaged.

1.3 Packaging

  1. Place the device in a carrier (e.g. a surgical instrument tray). Do not stack devices.

  2. Double wrap in accordance with ISO 11607-1.

1.4 Sterilization

Use a fractionated vacuum sterilization method in accordance with ISO 17665 and national recommendations. This method has been validated to achieve the desired sterility assurance level (SAL) of 10-6.

Phase 1 – Conditioning

A vacuum is generated up to four times. Steam is then injected to warm the device.

Phase 2 – Sterilization

The validated parameters for the fractionated pre-vacuum method are:

Phase 3 – Evacuation

The steam is released.

Phase 4 – Drying

Drying takes place under application of a vacuum.

After sterilization, cool the device uniformly at room temperature.

Sterilization validation was performed with an autoclave Hospithera CISA 6464LS Steam Sterilizer and MED-EL Surgical Instrument Tray (REF 41452) double wrapped in Halyard Sequential Sterilization Wrap H500. Drying was done over a period of at least 30 minutes.

1.5 Storage After Sterilization

Store the wrapped and sterilized device in accordance with the specifications provided by the manufacturer of the sterilization wrapping material.

Device: Positioning Jig

There are two methods for processing the Positioning Jig: disinfection or sterilization. Both are described below:

1 Processing - Sterilization approach

The processing consists of the following steps:

The device is a patient-matched device to be manufactured and processed during surgery at the point of use. Due to time constraints associated with the anesthesia duration, an automated cleaning process cannot be used, as it would extend the anesthesia time to an unacceptable length.

Warning

It is the responsibility of the user to use the validated processing recommended in these instructions for use and to use suitable equipment qualified and validated according to ISO 15883 series and other national recommendations.

The instructions provided in these instructions for use have been validated by the manufacturer of the medical device as being capable of preparing a medical device for use. It remains the responsibility of the processor to ensure that the processing, as actually performed, using equipment, materials, and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.

1.1 Manual Cleaning and Draining

Perform manual cleaning and drying:

  1. Wash with a cleaning solution (gigasept® pearls (2%)) at ambient temperature for at least 10 minutes. Ensure that all surfaces are covered by the cleaning solution. Drain cleaning solution from the device.

  2. Rinse with cold water 4 for at least 1 minute – drain

1.2 Inspection

Visually inspect the device for damages and remaining contamination. If there is remaining contamination, repeat cleaning process and inspection.

Caution

Discard the device if damaged.

1.3 Packaging

  1. Place the device in a carrier (e.g. a surgical instrument tray). Do not stack devices.

  2. Double wrap in accordance with ISO 11607-1.

  3. Label the packaging with a patient identifier to prevent mix-ups.

1.4 Sterilization

Use a fractionated vacuum sterilization method in accordance with ISO 17665 and national recommendations. This method has been validated to achieve the desired sterility assurance level (SAL) of 106.

Phase 1 – Conditioning

A vacuum is generated up to four times. Steam is then injected to warm the device.

Phase 2 – Sterilization

The validated parameters for the fractionated pre-vacuum method are:

Phase 3 – Evacuation

The steam is released.

Phase 4 – Drying

Drying takes place under application of a vacuum.

After sterilization, cool the device uniformly at room temperature.

Sterilization validation was performed with an autoclave Hospithera CISA 6464LS Steam Sterilizer and MED-EL Surgical Instrument Tray (REF 41452) double wrapped in Halyard Sequential Sterilization Wrap H500.

2 Processing - disinfection approach

The processing consists of the following steps:

The device is a patient-matched device to be manufactured and processed during surgery at the point of use. Due to time constraints associated with the anesthesia duration, an automated cleaning process cannot be used, as it would extend the anesthesia time to an unacceptable length.

Warning

It is the responsibility of the user to use the validated processing recommended in these instructions for use and to use suitable equipment.

The instructions provided in these instructions for use have been validated by the manufacturer of the medical device as being capable of preparing a medical device for use. It remains the responsibility of the processor to ensure that the processing, as actually performed, using equipment, materials, and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.

2.1 Manual Cleaning and Disinfection

Perform manual cleaning, disinfection, and drying:

  1. Prepare a cleaning/disinfection solution of Gigasept® Pearls (2%) in accordance with the manufacturer's instructions.

  2. Place the device in the cleaning/disinfection solution at ambient temperature for at least 15 minutes. Ensure that all surfaces are covered by the cleaning/disinfection solution. OtoJig recommends the use of the OtoJig disinfection bath with the included insert.

  3. Use sterile instruments/gloves to remove the device from the cleaning/disinfection solution. Drain the cleaning/disinfection solution from the device.

  4. Rinse with sterile water for at least 1 minute – drain.

2.2 Inspection

Visually inspect the device for damage and remaining contamination. If there is any remaining contamination, repeat steps 2.1 and 2.2.

Caution: Discard the device if damaged.

2.3 Packaging

  1. Place the device in a self-adhesive sterile bag.

  2. Seal the sterile bag with the adhesive sealer. Make sure that no air leaves the bag if it is pressed. Replace leaking bags.

  3. Label the carrier with a patient identifier to prevent mix-ups.

Devices: Mini-Stereotactic Frame and X-Ray Marker

1 Reprocessing

Reprocessing instructions according to ISO 17664-1.

Warning

It is the responsibility of the user to use the validated reprocessing recommended in these instructions for use and to use suitable equipment qualified and validated according to ISO 15883 series and other national recommendations.

The instructions provided in these instructions for use have been validated by the manufacturer of the medical device as being capable of preparing a medical device for use. It remains the responsibility of the processor to ensure that the reprocessing, as actually performed, using equipment, materials, and personnel in the reprocessing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.

1.1 Initial Treatment at the Point of Use

  1. Prepare a bath with demineralized water.

  2. Immediately after use, immerse the device in the bath to prevent drying.

The device must be reprocessed as soon as reasonably practical following use and not later than 24 hours after use.

Store and transport the device to the reprocessing site in a moist environment. Use a closed container in order to avoid contamination of the environment.

1.2 Preparation Before Automated Cleaning

Every component must be disconnected prior to cleaning, disinfection and sterilization.

Manual Pre-Cleaning

  1. Prepare a bath with demineralized water.

  2. Carefully and gently brush the device in the bath of demineralized water for at least 30 seconds until all visible soil is removed.

1.3 Automated Cleaning, Disinfection and Drying

Automated cleaning must be performed immediately after manual pre-cleaning.

  1. Prewash with cold water for at least 4 minutes – drain

  2. Wash with a detergent (Neodisher® MediClean forte (1%)) at +55 °C for at least 5 minutes – drain

  3. Rinse with cold water 3 for at least 1 minute – drain

  4. Disinfect with hot water 3 at +93 °C for at least 5 minutes – drain

  5. Dry with hot air at +100 °C for at least 20 minutes

Validation was performed with a washer/disinfector Belimed WD 290 IQ.

1.4 Inspection

Caution: Discard the device if damaged, worn, corroded or if the labeling is illegible.

Visually inspect the device for remaining soil in a well-lit area; no visible soil or moisture must be left on the device. If there is remaining debris, repeat the cleaning process and re-examine.

1.5 Packaging

  1. Place the device in a suitable carrier. Do not stack devices.

  2. Double wrap in accordance with ISO 11607-1.

1.6 Sterilization

Use a fractionated vacuum sterilization method in accordance with ISO 17665 and national recommendations. This method has been validated to achieve the desired sterility assurance level (SAL) of 106.

Phase 1 – Conditioning

A vacuum is generated up to four times. Steam is then injected to warm the device.

Phase 2 – Sterilization

The validated parameters for the fractionated pre-vacuum method are:

Phase 3 – Evacuation

The steam is released.

Phase 4 – Drying

Drying takes place under renewed application of a vacuum over a period of at least 30 minutes.

After sterilization, cool the device uniformly at room temperature.

Sterilization validation was performed with an autoclave Hospithera CISA 6464LS Steam Sterilizer and MED-EL Surgical Instrument Tray (MED-EL REF 41452) double wrapped in Halyard Sequential Sterilization Wrap H500.

1.7 Storage After Sterilization

Store the wrapped and sterilized device in accordance with the specifications provided by the manufacturer of the sterilization wrapping material.

Storage and transportation

Store all devices in a dry and clean environment, protected from direct sunlight, pests and extremes of temperature and humidity.

Store sterilized devices according to the (re-)processing instructions.

Store sterile delivered devices as indicated on the label of the sterile barrier system / packaging.

Use products in the order in which they are received (“first-in, first-out principle“), taking note of any expiration date on the label.

Only use the original packaging for transport as damage may occur otherwise. If the original packaging materials are no longer available, please contact OtoJig.

Lifetime of the Devices

The single-use (excl. Positioning Jig) devices are intended for short term use; their lifetime is limited to the duration of the procedure. The shelf life is limited to five years.

Positioning Jig: The lifetime is limited to the duration of the procedure. The shelf life is limited to 24 hours.

The reusable devices are intended for short term use; their lifetime is limited to 50 reprocessing cycles. The shelf life is limited to five years.

Disposal of the Devices

Dispose of in accordance with federal, state, and local regulations.

Dispose the devices Round Window Opener and Mini-Stereotactic Frame in a suitable container for sharp devices.

Please return used Positioning Jigs after cleaning. OtoJig may provide a discount for the next order.

OtoJig Surgical Procedure:

This procedure can be used for

Note

Activities related to the anesthesia and the application of drugs are not covered by the following steps and are to be performed according to the clinical routine.

Please consult the eIFU of conCIerge in latest version at https://otojig.com/IFU.

Optional screening: Pre-planning in conCIerge

If a pre-operative CT/CBCT volume scan is available for a potential patient, it is recommended to use conCIerge for screening to alleviate the intra-operative risk that no (safe) access tunnel can be planned for a given patient due to anatomic constraints.

Create new case in conCIerge

Check availability of devices

The following devices have to be available for creating a minimally invasive surgical tunnel in the temporal bone and to provide a covered access to the inner ear:

The following devices have to be available for performing a cochlear implantation:

Prepare the point of care production machine - Otomat

Prepare the point of care production machine (Otomat) according to the instructions for use of the Otomat.

Facial nerve monitoring

FacialNerveMonitoring

Setup facial nerve monitoring according to its IFU.

Caution

Facial nerve monitoring is mandatory.

Prepare patient

IMG_2343
IMG_2389

Skin incision and fixation of the first Anchor with a Bone Screw

IMG_2360
IMG_2368
IMG_2373

 

Fixation of the Mini-Stereotactic Frame to the first Anchor

IMG_2379

Fixation of the second Anchor and Optical Reference Marker



Mount X-Ray Marker

Caution

The contact surfaces of the X-Ray Marker with the MSF must be clean and free of debris to ensure accurate fit.

Warning

At this point, do not attempt to retighten the screws, reposition the MSF or apply excessive forces. The planning of the access path relies on the fact that the MSF remains at its position relative to the patient.

IMG_2406 IMG_2408 IMG_2410

Caution

Be careful not to damage your gloves when pressing the Jig Fastener pin into the shell.

IMG_2396

CT/CBCT scan

Import DICOM into conCIerge

Please consult the eIFU of conCIerge in latest version at https://otojig.com/IFU in order to:

An automated check for suitability of the scan is performed. If the scan is found not to be suitable, a new CT/CBCT scan can be performed.

Remove X-Ray Marker

Path planning

Please consult the eIFU of conCIerge in latest version at https://otojig.com/IFU and

Individualization of Positioning Jig

Mount Positioning Jig

IMG_2405
IMG_2403
IMG_2406 2
IMG_2419 2

Recommendation

The Positioning Jig may be hot after sterilization. Ensure that the Positioning Jig has reached ambient temperature before use, e.g. by cooling the Positioning Jig in a water bath.

Drilling — access tunnel

Important

Insertion and drilling should be performed while standing or at a similar level as the patient. One way to achieve this is by lowering the operating table.

OtoDrill 1

Continue drilling through the bone in ~2-3 second intervals. Retract the OtoDrill 1 from the drilled canal and apply cooling water into the canal between the intervals.

OtoDrill 2

 

 

Bony overhang removal - OtoDrill 3 🌀

If, in addition to the middle ear access, inner ear access is desired, continue as follows:

Implant bed and groove for CI 🌀

IMG_2464 2

Round window opening 🌀

IMG_2474

 

Removal of OtoJig devices

Covering the canal using the Insertion Tube 🌀

IMG_2487

Note

The Insertion Tube has a typical lumen of 1.27 mm and outside diameter of 1.33 mm. Devices larger than the lumen of the Insertion Tube cannot be used in the covered canal.

The Insertion Tube has a length of 31 mm and its inserted depth can be estimated by comparing the remaining non-inserted length with the planning in software conCIerge.

Aseptically remove the device from the sterile barrier system for presentation. Use the pull tab to remove the pouch and blister lid.

Caution

Do not apply large forces on the Insertion Tube as this may damage the device or tissue. Stop inserting the Insertion Tube when resistance is felt or the tube has reached the desired depth.

IMG_2490IMG_2500

IMG_2503IMG_2513

Glossary of symbols

SymbolTitleDescriptionReference         
Artboard3ManufacturerIndicates the medical device manufacturerISO 15223-1
Artboard6Date of manufactureIndicates the date when the medical device was manufacturedISO 15223-1
Artboard7Use-by dateIndicates the date after which the medical device is not to be usedISO 15223-1
Artboard8Batch codeIndicates the manufacturer’s batch code so that the batch or lot can be identifiedISO 15223-1
Artboard9Catalogue numberIndicates the manufacturer’s catalogue number so that the medical device can be identifiedISO 15223-1
Artboard2Country of manufactureTo identify the country of manufacture of productsISO 15223-1
Artboard11_3Sterilized using irradiationIndicates a medical device that has been sterilized using irradiationISO 15223-1
Artboard11_2Do not use if package is damaged and consult instructions for useIndicates that a medical device that should not be used if the package has been damaged or opened and that the user should consult the instructions for use for additional informationISO 15223-1
Artboard10Double sterile barrier systemIndicates two sterile barrier systemsISO 15223-1
Artboard4Single sterile barrier system with protective packaging outsideIndicates a single sterile barrier system with protective packaging outsideISO 15223-1
Artboard11Keep away from sunlightIndicates a medical device that needs protection from light sourcesISO 15223-1
Artboard12_3Keep dryIndicates a medical device that needs to be protected from moistureISO 15223-1
Artboard12Do not re-useIndicates a medical device that is intended for one single use onlyISO 15223-1
Artboard13_7Consult instructions for use or consult electronic instructions for useIndicates the need for the user to consult the instructions for useISO 15223-1
Artboard13_6CautionIndicates that caution is necessary when operating the device or control close to where the symbol is placed, or that the current situation needs operator awareness or operator action in order to avoid undesirable consequencesISO 15223-1
Artboard13_2Medical deviceIndicates the item is a medical deviceISO 15223-1
Artboard13_5Unique device identifierIndicates a carrier that contains unique device identifier informationISO 15223-1
CE markCE marking‘CE marking’ means a marking by which a manufacturer indicates that a device is in conformity with the applicable requirements set out in Regulation (EU) 2017/745 and other applicable Union harmonization legislation providing for its affixingRegulation (EU) 2017/745
onlyPrescription use onlyCaution: Federal law restricts this device to sale by or on the order of a physician21 CFR 801.109 (FDA)
Artboard13_4Health Industry Bar CodeThe HIBC emblem is a sign for optical marking linear and 2-dimensional bar codes for indicating that the code contains the standard HIBC structure.HIBC suggestion
🌀Steps necessary for inner ear access.If additionally to a minimally invasive access tunnel in the temporal bone, an access to the cochlear is desired, these steps and devices are required.Custom. Used in this IFU.

Reporting of serious incidents

Any serious incident that has occurred in relation to a device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established. Serious incident means any incident that directly or indirectly led, might have led or might lead to any of the following:

Use the following contact for reporting of serious incidents: reporting@otojig.com or call +49 511 53859999

Contact

Address:

https://otojig.com

Email: hello@otojig.com

 

 

The pictures provided in this material are for illustration purposes only and may not represent an actual device specification.

The trademarks used herein are trademarks of OtoJig GmbH, KLS Martin SE & Co. KG, Synthes GmbH, and MED-EL Elektromedizinische Geräte GmbH, respectively.

 

 


1 The names are used to simplify the surgical instructions. The drill bits "OtoDrill 1". "OtoDrill 2", and "OtoDrill 3" are used in the surgical procedure to clarify the order of usage of the drill bits.
2 Structures in the human temporal bone that could potentially be damaged include but are not limited to the inner ear including the round window membrane, the external ear canal wall, the tympanic membrane, the ossicles, the middle-ear muscles, blood vessels such as the sigmoid sinus, semi-circular canals, vestibule, the bony structures of the temporal bone, auditory nerve, vestibular nerve, chorda tympani, and facial nerve.
3 Used water shall meet the pharmacopendial limits of “Water for Injection”, at least with regards to bacterial endotoxins, i.e., 0.25 EU/mL.
4 The used water shall meet the national requirements for drinking water.